| Specifications | ||
|---|---|---|
| Power Supply | (EX12-1 Series) 20–28 V DC reverse polarity protected. (EX12-2 Series) 20–28 V DC reverse polarity protected or 24 V AC ± 10 % (21.6–26.4 V AC) ELECTRICAL CONNECTION) | |
Power consumption
| (24 V DC) 3.3 W, 90 mA, max. 130 mA | |
| Control unit | Microprocessor with 12-bit converter resolution | |
| Digital filter | Averaging in order to increase the EMC immunity | |
| Visual Indications | 3 LEDs for power, alarm and fault | |
| Analog output signal (active) | Proportional, overload and short-circuit proof, Max. load for UE > 20 V = 350 Ω and UE > 22 V = 500 Ω 4–20 mA = measuring range 3.3–4 mA = tolerable underrange 20–21.2 mA = tolerable overrange ≥ 21.2 mA= error overrange ≤ 2 mA = fault ≤ 1 mA = processor or voltage breakdown | |
| Serial Interface | Serial data bus | |
| Fault Relay (SPNC) | Max. 30 V AC/DC, 1 A | |
| Alarm Relay (SPDT) | Max. 30 V AC/DC, 1 A | |
| LCD (optional) | 2 x 16 characters, 3 status LEDs, 4 menu operating elements | |
| SENSOR ELEMENT | ||
| Measuring Principle | MPS™ | |
| Resolution | 0.1 % LEL | |
| Accuracy | ± 2 % LEL | |
| Zero-point variation | 0.5 % LEL | |
| Temperature range | -40 °C to 70 °C (-40 °F to 158 °F) | |
| Humidity range | 0–100 % RH non-condensing | |
| Pressure range | 80–120 kPa | |
| Storage temperature range | -40 °C to 60 °C (-40 °F to 140 °F) | |
| Life-time | 15+ years | |
| Calibration | 15 years | |
| Poisoning | Poisoning resistant | |
| Material | Aluminum pressure die-casting, epoxy coating | |
| Colour | RAL 7032 (pebble grey) Natural | |
| Dimensions housing (W x H x D) | Type 1 and 3: 125 x 167 x 83 mm (4.92 x 6.57 x 3.27 in) Type 2: 145 x 170 x 107 mm (5.71 x 6,70 x 4.21 in) Type Stainless: 145 x 170 x 107 mm | |
Weight: Type1, 2, 3 Type Stainless: | Approx. 1.3 kg (2.87 lb) / 1,6 kg (3,53 lb) Approx. 2.5 kg (5.51 lb) | |
Cable entry: (Type 1, 2, 3) Type: Stainless | 1x, 2x or 3x ¾ in. (Ansi B1.20.1) 2x NPT ¾“ (Ansi B1.20.1) | |
| Protection class | Housing protection IP66 to IP68 (depending on the cable glands used) | |
| Mounting | Wall mounting (sensor head downwards) | |
| Wire connection | Spring-type terminal, 0.08–2.5 mm² (AWG 28–14) | |
| Temperature Explosion protection With display | -40 °C to +60 °C (-40 to 140 °F) -20 °C to +60 °C (-4 °F to 140 °F) | |
| Pressure range | 80–120 kPa | |
| Air velocity | < 6 m/s | |
| EU Type Examination Certificate Electrical Explosion Protection ATEX | BVS 15 ATEX E 129 X EN IEC 60079-0:2018; EN 60079-1:2014 (DEKRA Testing and Certification GmbH) | |
| IECEx Type Examination Certificate Electrical Explosion Protection | IECEx BVS 16 0038X IEC 60079-0:2017; IEC 60079-1:2014-06 (DEKRA Testing and Certification GmbH) | |
| Type of Protection | (EX12-1 Zone 1) Ex db IIC T4 Gb -40 °C < Ta < +60 °C (EX12-2 Zone 2) Ex ec IIC T4 Gc -40 °C < Ta < +60 °C | |
| ATEX Marking | (EX12-1 Zone 1) II 2 G Ex db IIC T4 Gb (EX12-2 Zone 2) II 3 G Ex ec IIC T4 Gc | |
| EMC Test | Certificate PR 18 03 53984 001 EN 50270-2015 Interference immunity & emission: Type 2 (industrial sector) | |
EU Declaration of Conformity Electrical Explosion Protection | CE_PX2-1_Zone1 CE_PX2-2_Zone2 EN IEC | |
| Certificates (only housing) | ||
| Housing type 1 and type 3: FM Certificate of Compl. (3042541) Housing type stainless steel / type 2: FM Certif. of Conf. (FM18US0036U) | Class 3600, Class 3615, Class 3810, ANSI/NEMA 250. Explosion proof for Class I, Division 1, Groups A, B, C and D; dust-ignition-proof for Class II, Division 1, Groups E, F and G, Class III, hazardous (classified) locations, indoors and outdoors (type 4X). | |
| Housing type 1 and type 3: CSA Certif. of Compl. (2472857) Housing type stainless steel / type 2: CSA Certif. of Compl. (1717515) | Class 2258-02 PROCESS CONTROL EQUIPMENT – For Hazardous Loca tions Class 4418-02 OUTLET BOXES AND FITTINGS – For Hazardous Locations Class I, Div. 1, Groups A, B, C and D; Class II, Div. 1, Groups E, F and G, Class III, Div. 1; Type 4X | |
| Compliances of the sensor element | IEC EN 60079-1 / 60079-11 CSA 22.2 60079-1 / 60079-11 FM 3600 / 3610 ANSI/UL 913 JEDEC JS001-2017 EN 55011 IEC EN 61000-4-3 IEC EN 61000-4-8 IEC 60335-2-40 UL/CSA 60335-2-40 ASHRAE Standard 15 ASHRAE proposed Standard 15.2P JRA Standard 4068T:2016R | |
- Tunnel
- Food
- Beverage Dispensing
- Laboratory
- Climate
- Shipping
- Hydrogen
- Battery
- Biogas
- Agriculture
- Waste
- Process
- Gas storage
| Document | Download |
|---|---|
| Datasheet MPS LEL | Download |
| Datasheet Toxic & Oxygen | Download |
| Display Functions | Download |
| Manual | Download |
| Model Codes | Download |

Coming Soon!